    K Number
    K021123

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-04-26

    (18 days)

    Product Code
    Regulation Number
    866.2560
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BacT/ALERT® SN Culture Bottles are used with the BacT/ALERT Microbial Detection System in qualitative procedures for the recovery and detection of anaerobic and facultative anaerobic microorganisms (bacteria) from blood and other normally sterile body fluids.

    Device Description

    The BacT/ALERT SN Plastic Culture Bottle was developed for the same intended use as the current BacT/ALERT SN Glass Culture Bottle, to provide suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and normally sterile body fluids. An inoculated bottle is placed into the BacT/ALERT Microbial Detection Instruments where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT SN Bottle.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the BacT/ALERT SN (Plastic) Culture Bottle, focusing on acceptance criteria and the supporting study:

    Acceptance Criteria and Device Performance

    The provided document describes a substantial equivalence submission, which means the new device (BacT/ALERT SN Plastic Culture Bottle) is compared to a legally marketed predicate device (BacT/ALERT SN Glass Culture Bottle). The acceptance criteria are implicitly defined by demonstrating equivalent performance to the predicate device.

    Table 1: Acceptance Criteria and Reported Device Performance

    Performance Metric/CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (BacT/ALERT SN Plastic Culture Bottle)
    Microorganism RecoveryEquivalent recovery of low levels of 9 test microorganisms to the predicate device."Recovery of low levels of the 9 microorganisms included in the study" was equivalent to the predicate device.
    Detection TimesEquivalent detection times to the predicate device."Detection times were equivalent in both bottles."

    Study Information

    The type of study conducted is a comparative performance study against the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document states "Seeded studies were performed on 9 organisms." This indicates that nine specific microorganisms were tested. The number of replicates or individual tests per organism is not specified in this summary.
    • Data Provenance: The study involved "human blood," indicating the biological matrix used. The origin country of the human blood or where the study was conducted is not specified. The study is by nature prospective, as it involves direct experimentation with the new device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This type of study (microbial growth and detection) typically relies on laboratory-based microbiological techniques as the ground truth, rather than expert interpretation of images or clinical data. Therefore, no external experts were explicitly mentioned for establishing ground truth as one might see in imaging studies. The "ground truth" for positive growth and detection times would be determined by the instrument's (BacT/ALERT Microbial Detection Systems) readings and potentially confirmed by standard microbiological methods (e.g., subculturing). The qualifications of the laboratory personnel performing these tests are not specified.

    4. Adjudication Method:

    • None explicitly mentioned or applicable in the context of this performance study. The device's automated detection system and established microbiological methods would determine the outcome directly.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation and aims to measure the impact of AI assistance on human performance. This submission is for an automated microbial detection system, not an AI-assisted diagnostic tool for human readers.

    6. Standalone (Algorithm Only) Performance Study:

    • Yes, implicitly. The study directly measures the performance of the BacT/ALERT SN Plastic Culture Bottle in conjunction with the BacT/ALERT Microbial Detection System. Since the system (including the bottle) provides an automated output (detection of microbial growth), this can be considered a standalone performance assessment of the device and its associated technology without human intervention in the detection process itself.

    7. Type of Ground Truth Used:

    • The ground truth is based on direct microbiological observation/culturing and instrumental detection. The "ground truth" for microorganism recovery and detection times is established by the known inoculation of specific organisms and the automated detection by the BacT/ALERT Microbial Detection System, which is a well-established technology for this purpose.

    8. Sample Size for the Training Set:

    • Not applicable/not specified. This document describes a performance validation study for a device, not the development or training of a machine learning algorithm. Therefore, there is no "training set" in the context of AI/ML model development.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no training set for an AI/ML algorithm described, this question is not relevant to the provided documentation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K993421

    Validate with FDA (Live)

    Device Name
    BACT/ALERT SN
    Date Cleared
    1999-12-14

    (63 days)

    Product Code
    Regulation Number
    866.2560
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BacT/ALERT SN Culture Bottles are used with the BacT/ALERT Microbial Detection Systems in qualitative procedures for recovery and detection of anaerobic microorganisms (bacteria) from blood and other normally sterile body fluids.

    Device Description

    The BacT/ALERT SN Culture Bottle was developed for the same intended use as the current BacT/Alert Standard Anaerobic Culture Bottle, to provide suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and normally sterile body fluids. An inoculated bottle is placed into the BacT/ALERT Microbial Detection Instruments where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT SN Bottle.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BacT/ALERT SN Culture Bottle, based on the provided text:

    Acceptance Criteria and Device Performance

    The core of the acceptance criteria for the BacT/ALERT SN Culture Bottle revolved around demonstrating substantial equivalence to its predicate device, the BacT/Alert Standard Anaerobic Culture Bottle. This was primarily assessed by comparing the recovery and detection times of various microorganisms.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance (BacT/ALERT SN)
    Microorganism RecoveryEquivalent recovery of low levels of 11 specific microorganismsDemonstrated equivalent recovery of low levels of the 11 microorganisms included in the study compared to the predicate device.
    Detection TimesComparable detection times for the 11 specific microorganismsDetection times were comparable in both bottles (BacT/ALERT SN and predicate).
    Technological EquivalenceSame detection technologyUtilizes the same reflectance technology.
    Color Change MechanismColor change based on CO2 productionYes, exhibits color change based on CO2 production.
    Indicator MaterialSame indicator material as predicateYes, uses the same indicator material.
    Instrument CompatibilityCompatibility with BacT/ALERT Microbial Detection SystemsCompatible with BacT/ALERT Microbial Detection Systems.
    Sample SourceIntended for blood and body fluidsProcesses blood and body fluids.
    Target PopulationIntended for adult populationIntended for the adult population.

    Study Details

    This document describes a 510(k) Premarket Notification, which typically relies on demonstrating substantial equivalence rather than a full-scale clinical trial with acceptance criteria defined by specific clinical endpoints like sensitivity or specificity percentages. The study performed here is a comparative non-clinical study.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: "Seeded studies were performed on 11 organisms diluted in human blood and inoculated into the BacT/ALERT SN bottle and the BacT/Alert Standard Anaerobic bottle." The exact number of replicates or individual test instances for each organism is not specified.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA, suggesting the study was likely conducted in the U.S. or by a company with operations there.
      • Retrospective or Prospective: The description "Seeded studies were performed" implies a prospective experimental design, where samples were created and tested specifically for the study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This type of study (comparative performance of culture bottles) does not typically involve human experts establishing a "ground truth" in the same way an imaging AI algorithm might. The ground truth in this context is the confirmed presence and growth rate of the seeded microorganisms, which is measured by the BacT/ALERT system itself. The expertise would lie in the microbiological techniques used for organism preparation, inoculation, and confirmation, rather than diagnostic interpretation. The document does not specify the number or qualifications of these microbiological experts.

    4. Adjudication method for the test set:

    • No adjudication method is described, as the "outcome" is a direct measurement by the BacT/ALERT instrument (detection of growth and time to detection). The comparison is between the new device and the predicate device's performance, not human interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This was a non-clinical, in-vitro diagnostic (IVD) device study comparing the performance of two culture bottles. MRMC studies are typically used for imaging AI algorithms where human interpretation is a key component.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, effectively. The BacT/ALERT system (both with the new SN bottle and the predicate bottle) operates as an automated, standalone system for detecting microbial growth. The study assessed the performance of the new bottle within this automated context. There isn't a "human-in-the-loop" for interpreting the primary outcome of microbial growth detection; the instrument provides a positive/negative result and a time to detection.

    7. The type of ground truth used:

    • The ground truth was established by seeded microorganisms of known species and concentrations, diluted in human blood. The expected outcome was the growth of these specific microorganisms. The "truth" is the fact of microbial presence and their growth characteristics.

    8. The sample size for the training set:

    • This document does not describe a training set. This is not an AI/machine learning device that requires a training set. It's a medical device (culture bottle) that undergoes performance testing.

    9. How the ground truth for the training set was established:

    • As there is no training set for this device, this question is not applicable.